Infrared spectrum analysis can be roughly divided into two aspects: functional group qualitative analysis and structural analysis. Functional group qualitative analysis is to identify which groups a substance contains based on the characteristic group frequency of the infrared spectrum of the compound, thereby determining the category of the relevant compound. Structural analysis requires the infrared spectrum of the compound to infer the chemical structure of the relevant compound. For the structural analysis of a new compound, it is necessary to cooperate with ultraviolet absorption spectrum, nuclear magnetic resonance spectrum and mass spectrum.
1. Sample preparation
1) Solid samples
Solid samples are usually prepared by tableting. Grind the sample into fine powder and add dry KBr powder. After grinding, put it into a tablet press and press it on a tablet press while evacuating air to make a transparent sheet of a certain diameter and thickness. KBr has no characteristic absorption in the mid-infrared region, so the KBr sheet containing the sample is placed in the optical path of the instrument to measure the infrared absorption spectrum of the sample.
2) Liquid samples
Liquid samples with low volatility can be directly applied to the KBr sheet (smear method), or dropped between two KBr sheets to form a liquid film (liquid film method) and placed in the optical path for measurement.
Samples with high volatility can be injected into a closed liquid pool for measurement (liquid pool method). For some liquid samples with strong infrared absorption, solutions can be made to reduce the concentration; solid samples and gaseous samples can also be prepared into solutions using appropriate solvents and injected into a liquid pool for measurement. In addition to having sufficient solubility for the sample, the selected solvent should also have no strong absorption in the measured spectral region, no erosion of the salt window, and no strong solvation effect on the sample. Commonly used solvents include CCl4 (4000~1350 cm-1) and CS2 (1350~600 cm-1). CCl4 has a slight interference near 1580 cm-1.
3) Gaseous samples
Gaseous samples can be measured directly using a gas absorption cell or prepared as a solution using a liquid absorption cell.

2. Infrared absorption spectrum analysis
General process of infrared spectrum qualitative analysis:
1) Purification of samples
The purification of samples mainly includes separation and drying processes. The samples are purified by various separation methods such as fractionation, extraction, recrystallization, chromatography, etc., and dried to obtain dry pure substances. Impure samples will bring difficulties to the analysis of the spectrum.
2) Understand the source, properties and other experimental data of the sample
The source of the sample, elemental analysis results, relative molecular mass, melting point, boiling point, solubility and other properties, ultraviolet spectrum, nuclear magnetic resonance spectrum, mass spectrum and other data may be of great help to the analysis of infrared spectrum.
Understanding the source of the sample can narrow the scope of structural speculation. According to the elemental analysis value of the sample and the molecular formula obtained from the relative molecular mass, the degree of unsaturation can be calculated, and then it can be estimated whether the molecular structure contains double bonds, triple bonds and aromatic rings, and the rationality of the spectral analysis results can be verified. Properties such as boiling point and melting point can be used as circumstantial evidence for spectral analysis.
3) Prepare the sample using appropriate methods and record the infrared absorption spectrum.
4) Spectrum analysis
There is no definite rule for the analysis of the spectrum so far.
First, start the analysis from the first strong peak in the characteristic area. The absorption peak in the characteristic area has little interference and is easy to identify. According to the clues that a certain group in the characteristic area may exist, further verification is carried out in the fingerprint area to find the combination mode of the group with other groups.
Then, according to the peak position of the absorption peak to be checked, the origin and possible attribution of the absorption peak are preliminarily understood, and then its related peaks are found. The so-called related peaks refer to a group of interdependent characteristic peaks that appear in a functional group. In some cases, due to overlap with other peaks or the peak intensity is too weak, not all related peaks can be observed, but the related peaks must be found to confirm the existence of the functional group. It is an important principle to use a group of related peaks to identify the existence of a functional group.
Finally, the elimination method is used to deny the existence of certain functional groups in order to gradually narrow the scope.
5) Compare with standard spectra
In infrared spectroscopy, whether it is the verification of known substances or the identification of unknown substances, it is often necessary to use the spectra of pure substances for verification. These standard spectra can be made by yourself under the same conditions with pure substances, or you can refer to relevant standard spectrum collections. When comparing, you should pay attention to the following:
i. The aggregation state and sample preparation method of the test object and the standard spectrum should be consistent.
ii. The fingerprint area should be carefully compared. Because the spectral bands in the fingerprint area are very sensitive to slight changes in structure, slight changes in structure can lead to differences in the spectral bands in the fingerprint area.
Please kindly note that our products are for research use only.